Background Image
Table of Contents Table of Contents
Previous Page  19 / 302 Next Page
Information
Show Menu
Previous Page 19 / 302 Next Page
Page Background

Analysis of Early Eluting Pesticides in a

C18-Type Column Using a Divert Valve and

LC-MS/MS

Eugene Kokkalis

1

, Anna Rentinioti

1

, Kostas Tsarhopoulos

2

, Frans Schoutsen

3

1

Engene SA, Athens, Greece;

2

RigasLabs SA, Thessaloniki, Greece;

3

Thermo Fisher Scientific, Breda,

The Netherlands

Application Note

572

Key Words

TSQ Quantum Access MAX, Divert Valve, Split Peaks, Reversed-Phase

Liquid Chromatography, Pesticides

Goal

To demonstrate the ability to override the solvent effects from a sample

extract using gradient solvents with liquid chromatography. Additionally, to

increase injection volume without overloading the column.

Introduction

Many pesticide analyses are based on the QuEChERS

extraction method, which uses acetonitrile (ACN) in the

final extraction step. However, injecting a solvent stronger

than the HPLC mobile phase can cause peak shape problems,

such as peak splitting or broadening, especially for the

early eluting analytes (low capacity factor, k). The

common practice is to exchange the solvent of the final

extraction step for one similar to the mobile phase, for

example methanol / water, but this procedure is laborious

and can lead to analyte losses.

There are several possible causes of peak splitting or

broadening. This study presents the peak shape differences

between acetonitrile and methanol / water [1:1 v/v]

solutions due to the interaction of gradient and sample

solvent, as indicated in Figure 1. The lowest detection

limit is achieved when an analyte is in as compact a band

as possible within the flow stream of mobile phase and

with larger injection volumes. However, this is limited by

maximum loop volume and column capacity.

Mobile phase composition and the use of a divert valve

have been evaluated for the analysis of seven selected

pesticides in acetonitrile solutions (Table 1). The sample

solutions were chosen to represent both low and high

analyte levels for compounds that elute either early or

middle-early from a C18 column. Performance was

evaluated in terms of linearity (injection volume range

1–8 µL), robustness (RSD), and sensitivity as measured by

signal-to-noise ratio (S/N) and peak area reproducibility.

Figure 1. Chromatograms of 5 µL injections of acephate,

omethoate, oxamyl, methomyl, pymetrozin, and monocrotophos

in 50 µg/L acetonitrile (A) and methanol / water [1:1 v/v] solution

(B), with no divert valve used

A

B