Figure 3: Overlay of 20 repeated injections of Cr (VI) and Cr (III)

Quantification of Cr (III) and Cr (VI) in tap water

Potable water was collected locally and analyzed using the

proposed method. As can be seen from the chromatogram

in Figure 4, only trace amounts of Cr (VI), at a retention

time of ~40 s, could be detected in this sample. After

external calibration, the amount of Cr (VI) observed was

found to be 42.5 ± 1 pg/g. As an additional proof that the

detected peak corresponds to Cr and is not affected by

possibly co-eluting compounds causing spectral

interferences (e.g. chlorine or carbon based polyatomic

species), the isotope ratio

52

Cr

+

/

53

Cr

+

was calculated and

corresponds well to the theoretical value of 8.81).

Conclusions

Through the combination of the ICS-5000 ion

chromatography system with the iCAP Qc ICP-MS, a

sensitive, robust method for the speciation analysis of

trace levels of Cr (III) and Cr (VI) in natural waters has

been developed. The method developed enables fast and

reliable speciation analysis of both Cr (III) and Cr (VI)

species in water samples without prior incubation steps

and with high purity nitric acid as mobile phase. The

short, but highly efficient Dionex AG-7 column, provides

complete separation of both species in under 150 s,

enabling high sample throughput for the routine analysis

of water samples.

thermoscientific.com

© 2012 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organisation. All other trademarks are the property of

Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries.

Please consult your local sales representative for details.

Africa-Other

+27 11 570 1840

Australia

+61 2 8844 9500

Austria

+43 1 333 50 34 0

Belgium

+32 53 73 42 41

Canada

+1 800 530 8447

China

+86 10 8419 3588

Denmark

+45 70 23 62 60

Europe-Other

+43 1 333 50 34 0

Finland /Norway/Sweden

+46 8 556 468 00

France

+33 1 60 92 48 00

Germany

+49 6103 408 1014

India

+91 22 6742 9434

Italy

+39 02 950 591

Japan

+81 45 453 9100

Latin America

+1 608 276 5659

Middle East

+43 1 333 50 34 0

Netherlands

+31 76 579 55 55

South Africa

+27 11 570 1840

Spain

+34 914 845 965

Switzerland

+41 61 716 77 00

UK

+44 1442 233555

USA

+1 800 532 4752

AN43098_E 03/12C

Application Note 43098

The new flatapole cell technology introduced in the iCAP

Q ICP-MS provides interference-free detection of the

52

Cr

and

53

Cr ions. Sub-ppt detection limits are achievable due

to the completely metal free pathway of the ICS-5000 and

the high instrumental sensitivity offered by the iCAP Q’s

He KED mode.

References

1. Séby, F., Charles, S., Gagean, M., Garraud, H., Donard,

O. F. X., J. Anal. At. Spectrom. 18 (2003), 1386-1390

2. Xing, L., Beauchemin, D., J. Anal. At. Spectrom. 25

(2010), 1046-1055

3. Dionex homepage

(http://www.dionex.com/en-us/

products/columns/ic-rfic/specialty-packed/ionpac-as7/

lp-73274.html)

Chemicals Used in this Note

Chemical

Fisher Scientific

Catalogue Number

IonPac AG-7 Guard Column (2 x 50 mm)

063099

Fisher Optima grade nitric acid

A467-500

For more information please contact your local Fisher

Scientific organization and/or visit:

www.fishersci.com

or

www.acros.com

Thermo Fisher Scientific (Bremen) GmbH

Management System Registered to ISO 9001:2008

Figure 4: 52Cr and 53Cr chromatograms of a locally source potable water.