Table 2. EQuan MAX chromatography conditions used

Table 3. Exactive Orbitrap MS operating conditions

Parameter

Setting

Scan range

m/z

150–1100

Resolving power

50,000 (FWHM at

m/z

200)

Polarity

Positive

Measured

m/z

995.5543 MC-LR

519.7898 MC-RR

1045.5344 MC-YR

825.4501 Nodularin

Ionization source

Electrospray

Spray voltage

4000 V

Capillary temperature

340 °C

Capillary voltage

37 V

Tube lens voltage

85 V

Skimmer voltage

22 V

Results and Discussion

High-Resolution Mass Spectra of Toxins

The standards were prepared at a concentration of 1 ng/mL

each and injected using a syringe pump to observe the

mass spectra. The molecular ion and carbon isotope

spectra of microcystin-LR, -RR, -YR, and nodularin are

shown in Figure 2a. Four carbon isotopes were observed

for most compounds. Using this isotopic pattern, it was

possible to match the experimentally recorded carbon

isotopic distribution ratios to the theoretical isotopic ratio

to provide confirmation of the toxin using the analysis

software. Meanwhile, molecular ions were observed in

nodularin at

m/z

825 and the isotopic pattern was

confirmed (Figure 2b).

Pump 1

Pump 2

Hypersil GOLD aQ

(preconcentration column)

Hypersil GOLD

(analytical column)

Time

%A

%B µL/min Time

%A

%B µL/min

0.00

98

2

1000

0.00

98

2

400

1.01

98

2

1000

1.00

98

2

400

1.20

98

2

100

2.00

2

98

400

3.50

98

2

100

3.50

2

98

400

4.00

98

2

1000

3.51

98

2

400

4.00

98

2

400

Mobile phase A: 0.1% formic acid in water

B: acetonitrile

Mobile phase A: 0.1% formic acid in water

B: acetonitrile

Column temperature: Ambient

Injection volume: 1000 µL