High Resolution LC-MS for Screening and Quantitative Analysis of Antibiotics in Drinking Water Using an Orbitrap and Online Sample Preparation

Conclusion

This poster demonstrates:

Online pre-concentration and extraction for 1mL injections of antibiotics at the ppt

level.

The quantitation of HR/AM data using TraceFinder software from the Exactive

Plus Orbitrap instrument.

Spectral confirmation of MS

2

spectrum collected in the same data file as the

quantitation data using ExactFinder software.

Ability to quantitate and confirm samples in the same analytical run for antibiotics

in water samples.

Acknowledgements

We would like to thank Paul Yang from the Ontario Ministry for the Environment for

supplying the antibiotic standards for this analysis.

All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to

encourage use of these products in any manners that might infringe the intellectual property rights of others.

ples, spectral

rformed using

fter acquisition. The

Food Safety and

ectra collected on

eters are

e 4. This

ard corresponding

rence spectrum.

reference

ntific LTQ Velos Pro

less background,

eless, the two

d molecular ion at

the possibility of

ems.

sses, LOQs and

)

% RSD at LOQ

8.90

14.30

9.90

13.50

4.20

7.30

4.80

4.90

3.45

6.30

6.60

3.60

13.10

FIGURE 4. Spectral comparision of the MS

2

spectrum of the antibiotic

trimethoprim obtained at a concentration of 80 pg/mL. The library reference

spectrum is the top spectrum, the lower spectrum is from the sample. The

comparison was performed with ExactFinder software.