Lysergic Acid Diethylamide (LSD) in Serum, Plasma or

Whole Blood for GC or GC/MS Confirmations

Using 200mg 10mL HyperSep Verify-CX Extraction Column

(Part Number: 60108-742)

Prepare Sample

To 1mL of blood, plasma or serum, add 4mL of DI H

2

O

and internal standard*

Mix/vortex and let stand for 5 minutes

Centrifuge for 10 minutes at 2,000rpm and discard pellet

Add 2mL of 100mM phosphate buffer (pH 6.0)

Mix/vortex

Sample pH should be 6.0±0.5

Adjust pH accordingly with 100mM monobasic or dibasic

sodium phosphate

Condition Verify-CX Extraction Column

3mL of CH

3

OH then aspirate

3mL of DI H

2

O then aspirate

1mL of 100mM phosphate buffer (pH 6.0)

NOTE:

Aspirate at <3"Hg to prevent sorbent drying

Apply Sample

Load sample at 1 to 2mL/minute

Wash Column

3mL of DI H

2

O

3mL of 100mM acetic acid

3mL CH

3

OH

Dry column (5 minutes at >10"Hg)

Elute LSD

3mL of CH

2

Cl

2

/IPA/NH

4

OH (78:20:2)

Collect eluate at 1 to 2mL/min using minimal vacuum

NOTE:

Prepare elution solvent fresh daily; add IPA/NH

4

OH mix,

then add CH

2

Cl

2

(pH 11-12)

Dry Eluate

Evaporate to dryness at <40°C

Derivatize

Add 20µL ethyl acetate and 20µL BSTFA

(with 1% TMCS)***

Overlay with N

2

and cap

Mix/vortex

React for 30 minutes at 70°C

Remove from heat source to cool

NOTE:

Do not evaporate BSTFA solution

Quantitate

Inject 1 to 2µL onto gas chromatograph

For mass spectrometry, monitor the following ions:

Compound

Primary Ion** Secondary

Tertiary

D

3

-LSD-TMS*

298

296

271

LSD-TMS

395

293

268

* Suggested internal standard

** Quantitation ion

*** Part number TS-38831 (10 x 1mL ampules)

Recommended GC Columns

Part Number

TRACE TR-DoA5 Column, 15m x 0.25mm x 0.25µm

26AF130P

TRACE TR-DoA35 Column, 15m x 0.20mm x 0.33µm

26AC497P

100