Background Image
Previous Page  98 / 144 Next Page
Information
Show Menu
Previous Page 98 / 144 Next Page
Page Background

98

Forensic

Applications

Gamma-Hydroxybutyrate (GHB) in Urine without Conversion to

Gamma-Butrylactone (GBL)

Using 200mg 6mL HyperSep Retain-AX Extraction Column

(Part Number: 60107-412)

Sample Preparation

200µL of urine

Add internal standard* and 100µL of 100mM phosphate

buffer (pH 6.0)

Mix/vortex

Condition HyperSep Retain AX Extraction Column

1 x 3mL CH

3

OH

1 x 3mL DI H

2

O

1 x 1mL 100mM phosphate buffer (pH= 6.0)

Apply Sample

Place test tubes into vacuum manifold for collection

The sample loading and wash are both collected

Decant sample onto column. Aspirate at ~1"Hg

Wash Column

Add 1mL of CH

3

OH/NH

4

OH (99:1) to original sample

test tube

Mix/vortex

Decant wash onto column

Evaporate to dryness at 60°C using a stream of air or N

2

Sample Clean-Up

Add 200µL of dimethylformamide

Add 1mL of hexane saturated with dimethylformamide

Mix by inversion for 5 minutes

Centrifuge at 3,000rpm for 5 minutes

Transfer lower dimethylformamide layer to a clean test tube

Evaporate to dryness at <50°C using a stream of air or N

2

Derivatize

Add 100µL ethyl acetate and 100µL BSTFA

(with 1% TMCS)**

Mix/vortex

Analysis

Inject 1 to 2µL onto GC/MS

For mass spectrometry monitor the following ions:

Compound

Primary Ion *

Secondary

Tertiary

GHB-D6-di-TMS*

239

240

241

GHB-di-TMS

233

234

235

* Suggested internal standard for GC/MS: GHB-D6

** Part number TS-38831

Recommended GC Column

Part Number

TraceGOLD TG-35MS 30m x 0.25m x 0.25µm

26094-1420