Results and Discussion
LC-MS/MS is a highly selective technique for analyzing
drugs from biological matrices. As shown in Figure 1,
samples are loaded to the extraction column via an
autosampler and LC pump combination. After a short
washing time, the central valve is rotated 60° to the injec-
tion position, which allows a second pump to elute the
analytes from the extraction column onto the analytical
column and into the FAIMS-enabled mass spectrometer. In
cases where background or co-eluting interferences appear,
the limiting factor is selectivity.
FAIMS is a unique selectivity enhancing tool for LC-
MS. With FAIMS, gas-phase ions are purified after LC
analysis but before they are mass analyzed. The waveform
shown in Figure 2 separates the gas-phase ions as they are
transferred into the high vacuum region of the MS. The
interference ions in orange are filtered out from the ion
beam, while the analyte ions in blue pass into the MS.
The basic experiment that establishes FAIMS condi-
tions is shown in Figure 3. While infusing a reference
standard of testosterone into the mobile phase, the CV is
scanned over a specific range. The black trace appears for
the transition due to testosterone, and a maximum signal
for testosterone appears at CV –12.5V. To ensure that
mobile phase components with the same transition will
be eliminated, stop the infusion of the reference standard
and repeat the CV scan over the same specified range.
The red trace indicates that chemical background in the
mobile phase emerges from FAIMS not at the CV for
testosterone, but rather in a broad range between CV –15
and –25V. For subsequent LC-FAIMS-MS/MS assays, the
CV set to –12.5V will exclude mobile phase contributions
to the analysis of testosterone. Other endogenous compo-
nents present in the human serum samples may also be
excluded if their FAIMS behavior is different from that of
testosterone.
Regression analysis based on a linear model with
1/[concentration]
2
weighting was used. The average
accuracy, as deviation from theoretical is less than 5% at
all concentrations. The precision of the standards, as
relative standard deviation (%RSD), is less than 17% at
the lower limit of quantitation and less than 11% at all
other concentrations.
Despite the excellent performance of LC at cleaning
up samples, many interferences are still present from the
matrix. Other extraction techniques might remove these
Pump
Pump
Pump
Pump
Waste
Waste
Autosampler
Autosampler
Analytical Column
Analytical Column
FAIMS-enabled
MS/MS
FAIMS-enabled
MS/MS
Extraction
Column
Extraction
Column
Figure 1: Schematic diagram of LC-FAIMS-MS/MS system
Load sample to extraction column
Inject sample to analytical
column for FAIMS-MS/MS
analysis
1...,46,47,48,49,50,51,52,53,54,55 57,58,59,60,61,62,63,64,65,66,...374