DHEA, Testosterone and Epitestosterone in Urine

for GC or GC/MS Confirmations

Using 200mg 10mL HyperSep Verify-CX Extraction Column

(Part Number: 60108-742)

Prepare Sample

Pipette 5mL of urine into borosilicate glass test tubes

Add internal standard* and adjust the sample pH to

5.5 to 6.5 using concentrated monobasic or dibasic

sodium phosphate

Mix sample

Centrifuge samples at 3,000rpm for 5 minutes

Condition Verify-CX Extraction Column

3mL of CH

3

OH

3mL of DI H

2

O

1mL of 100mM phosphate buffer (pH 6.0)

Apply Sample

Pour supernatant onto column

Allow to flow via gravity

Wash Column

3mL of DI H

2

O

Dry column (10 minutes at >10"Hg)

Elute Steroids

3mL of CH

3

OH

Collect eluate at 1 to 2mL/min

Enzymatic Hydrolysis

Dry eluate under a stream of nitrogen; add 2mL of 0.2M

phosphate buffer (pH 7.0) and 250 units of

β

-glucuronidase

Mix/vortex and allow to incubate at 50°C for 1 hour

Cool sample, cap and adjust the pH to 10-11 using a

1:1 mixture of NaHCO

3

/Na

2

CO

3

Additional Clean-up

Add 5mL of n-butyl chloride to each sample. Cap the

tubes and shake vigorously for 10 minutes and then

centrifuge at 3,000rpm for 5 minutes. Transfer the

organic layer to a clean test tube and dry under a

stream of nitrogen. Place the dried sample in a desiccator

and further dry under vacuum for 30 minutes.

Derivatize

Add 50µL of MSTFA***/NH

4

/dithioerythritol (1000:2:5,

v/w/w) and incubate at 70°C for 20 minutes

Centrifuge sample at 3,000rpm for 1 minute and transfer

directly to GC injector vials

Quantitate

Inject 1 to 2µL onto the gas chromatograph

For mass spectrometry, monitor the following ions:

Compound

Primary Ion** Secondary

Testosterone

432

417

Epitestosterone

432

417

DHEA

432

417

16

α

hydroxyl-testosterone*

520

259

* Suggested internal standard at 20ng/mL

** Quantitation ion

*** Part number TS-48910 (10 x 1mL ampules)

Recommended GC Columns

Part Number

TRACE TR-DoA5 Column, 15m x 0.25mm x 0.25µm

26AF130P

TRACE TR-DoA35 Column, 15m x 0.20mm x 0.33µm

26AC497P