Background Image
Previous Page  72 / 144 Next Page
Information
Show Menu
Previous Page 72 / 144 Next Page
Page Background

72

Forensic

Applications

Blood GHB Extraction

Using 200mg 3mL HyperSep Retain-AX Extraction Column

(Part Number: 60107-404)

Sample Preparation

To 1mL blood sample add internal standard* and 0.5mL

of 100mM phosphate buffer (pH 6.0)

Mix/vortex

Rock for 10 minutes

Centrifuge for 10 minutes at 2700rpm

Condition HyperSep Retain-AX Extraction Column

1 x 3mL CH

3

OH

1 x 3mL DI H

2

O

1 x 1mL 100mM phosphate buffer (pH 6.0)

NOTE:

Aspirate at <3"Hg to prevent sorbent drying

Apply Sample

Place centrifuge tubes into vacuum manifold for collection

The sample loading is collected

Decant sample onto column

Aspirate at about 1"Hg

After the sample is off the columns apply full vacuum for

about 15 seconds to remove any residual blood

Elute

Remove centrifuge tubes, set aside

Place clean centrifuge tubes into vacuum manifold for

collection

1 x 2mL of CH

3

OH/NH

4

OH (99:1)

Aspirate at about 1"Hg

Concentrate

Remove test tubes from vacuum manifold

Mix/vortex the sample prior to concentrating

Evaporate to dryness at 60°C using a stream of nitrogen

Sample Clean-up

Add 200µL of dimethylformamide

Add 1mL of hexane saturated with dimethylformamide

Rock for 5 minutes

Centrifuge at 5 minutes at 2700rpm

Transfer lower dimethylformamide layer to a clean test tube

Evaporate to dryness at 50°C using a stream of air or

nitrogen

Derivatize

Add 25µL ethyl acetate and 25µL BSTFA

(with 1% TMCS**)

Mix/vortex

Heat at 70°C for 30 minutes

Analysis

Inject a 1 to 2µL of the sample onto GC/MS

Compound

Primary Ion *

Secondary

Tertiary

GHB-D6-di-TMS*

239

240

241

GHB-di-TMS

233

234

235

* Suggested internal standard for GC/MS: D6-GHB

** Part number TS-38831

Recommended GC Column

Part Number

TraceGOLD TG-5SilMS 30m x 0.25m x 0.25µm

26096-1420